3.4 Device Message Details
3-245
3
Message Details
[:SENSe][:WAVelength]:STOP
Function
This command sets and queries the stop wavelength.
Syntax
[:SENSe][:WAVelength]:STOP <numeric>[<unit_m>]
[:SENSe][:WAVelength]:STOP?
Range 600.0 to 1800.0 nm (Stop wavelength)
Resolution 0.1 nm
If a suffix is omitted, the unit will be m.
Response Data
<numeric>
<numeric>: Stop wavelength (m)
+6.00000000E007 to +1.80000000E006
Specify larger value than Start wavelength.
Example of Use
:SENSe:WAVelength:STOP 1562.8NM
:SENSe:WAVelength:STOP?
>+1.56280000E-006