3.4 Device Message Details
3-113
3
:CALCulate:PARameter[:CATegory]:LD:AREA
Function
This command sets and queries the Noise Parameter Area Type for the
LD Module application.
Syntax
:CALCulate:PARameter[:CATegory]:LD:AREA CH|USER
:CALCulate:PARameter[:CATegory]:LD:AREA?
CH: Sets Area Type to Channel
USER: Sets Area Type to User Specify
Response Data
CH|USER
Example of Use
:CALCulate:PARameter:CATegory:LD:AREA USER
:CALCulate:PARameter:CATegory:LD:AREA?
>USER
:CALCulate:PARameter[:CATegory]:LD:K
Function
This command sets and queries the magnification for standard deviation
in the LD Module application.
Syntax
:CALCulate:PARameter[:CATegory]:LD:K <numeric>
:CALCulate:PARameter[:CATegory]:LD:K?
<numeric>: k Standard deviation multiplier
Range 1.00 to 10.00
Resolution 0.01
Response Data
<numeric>
<numeric>: k Standard deviation multiplier
+1.00000000E+000 to +1.00000000E+001
Example of Use
:CALCulate:PARameter:CATegory:LD:K 6.07
:CALCulate:PARameter:CATegory:LD:K?
>+6.07000000E+000