Chapter 3 Message Details
3-240
[:SENSe]:SWEep:POINts
Function
This command sets and queries the number of sampling points.
Syntax
[:SENSe]:SWEep:POINts <numeric>
[:SENSe]:SWEep:POINts?
Specify any of the following to <numeric>.
51, 101, 251, 501, 1001, 2001, 5001, 10001, 20001, 50001
Response Data
51|101|251|501|1001|2001|5001|10001|20001|50001
Example of Use
:SENSe:SWEep:POINts 1001
:SENSe:SWEep:POINts?
>1001
:SENSe:SWEep:TIME:INTerval
Function
This command sets and queries the time interval of the sweep start.
Syntax
:SENSe:SWEep:TIME:INTerval <numeric>[S]
:SENSe:SWEep:TIME:INTerval?
Range 0 to 5940 s
Resolution 1 s
If a suffix is omitted, the unit will be s.
Response Data
<numeric>
<numeric>: Time interval (s) 0 to 5940
Example of Use
:SENSe:SWEep:TIME:INTerval 60S
:SENSe:SWEep:TIME:INTerval?
>60