Chapter 3 Message Details
3-258
:TRACe[:DATA][:Y]:DCA|DCB|DCC|DCD|DCE|DCF|DCG|DCH|DCI|DCJ?
Function
This command queries wavelength and sampling points of the trace.
Specify a trace (A to J) by specifying a parameter value from DCA to
DCJ.
Syntax
:TRACe[:DATA][:Y]:DCA|DCB|DCC|DCD|DCE|DCF|DCG|DCH|DCI|DC
J?
Response Data
<numeric>,<numeric>,<switch>
No.
Parameter Type
Range
Description
1
<numeric>
+6.00000000E–007 to
+1.75000000E–006
Start
wavelength (m)
2
<numeric>
+6.00000000E–007 to
+1.80000000E–006
Stop wavelength
(m)
3
<switch>
51|101|251|501|1001|2001|
5001|10001|20001|50001
Sampling points
Example of Use
:TRACe:DATA:Y:DCA?
>+1.55239000E+003,+1.55739000E+003,2001