EasyManua.ls Logo

Anritsu MS9740A - Trace[:Data][:Y]:Dca|Dcb|DCC|DCD|Dce|Dcf|Dcg|Dch|DCI|Dcj

Anritsu MS9740A
320 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Message Details
3-258
:TRACe[:DATA][:Y]:DCA|DCB|DCC|DCD|DCE|DCF|DCG|DCH|DCI|DCJ?
Function
This command queries wavelength and sampling points of the trace.
Specify a trace (A to J) by specifying a parameter value from DCA to
DCJ.
Syntax
:TRACe[:DATA][:Y]:DCA|DCB|DCC|DCD|DCE|DCF|DCG|DCH|DCI|DC
J?
Response Data
<numeric>,<numeric>,<switch>
No.
Parameter Type
Range
Description
1
<numeric>
+6.00000000E007 to
+1.75000000E006
Start
wavelength (m)
2
<numeric>
+6.00000000E007 to
+1.80000000E006
Stop wavelength
(m)
3
<switch>
51|101|251|501|1001|2001|
5001|10001|20001|50001
Sampling points
Example of Use
:TRACe:DATA:Y:DCA?
>+1.55239000E+003,+1.55739000E+003,2001

Table of Contents

Related product manuals