Chapter 3 Message Details
3-90
:CALCulate:MARKer[1|2|3|4]:X:FREQuency
Function
This command sets the frequncy of the trace marker and displays the
trace marker.
Syntax
:CALCulate:MARKer[1|2|3|4]:X:FREQuency
<numeric>[<unit_h>]
The process will not change regardless of whether a number [1|2|3|4] is
specified or omitted.
<numeric>: Frequncy of the trace marker
Range From the start frequency to the stop frequency
Resolution 0.01 GHz
If a suffix is omitted, the unit will be Hz.
If there is no data at the specified frequency point, the trace marker is
displayed at the closest data point to the specified frequency.
Example of Use
:CALCulate:MARKer:X:FREQuency 193.8THZ
:CALCulate:MARKer:X?
>+1.93800000E+014