3.4 Device Message Details
3-139
3
Message Details
:CALCulate:PARameter[:CATegory]:NF:PARameter
Function
This command sets the measurement parameters for the Optical AMP
(WDM) application.
This command queries the measurement parameters for the Optical AMP
(WDM) application.
Syntax
:CALCulate:PARameter[:CATegory]:NF:PARameter
<switch>,<switch>,<numeric>,<numeric>,<numeric>,<switch>
,<switch>
:CALCulate:PARameter[:CATegory]:NF:PARameter?
No.
Parameter type
Range
Description
1
<switch>
0|1
0:NF (S-ASE)
1:NF (Total)
2
<switch>
0|1|2
0: ISS Method (IEC)
1: ISS Method (Advanced)
2: Off
3
<numeric>
10.00 to 10.00 dB
Res. 0.01 dB
Pin Loss(Offset)
Loss correction factor for signal
level
4
<numeric>
10.00 to 10.00 dB
Res. 0.01 dB
Pout Loss(Offset)
Loss correction factor for optical
level output
5
<numeric>
0.100 to 10.000
Res. 0.001
NF Calibration
Correction factor for noise figure
6
<switch>
0|1
0: Actual Resolution (Measured)
1: Actual Resolution (Initial)
7
<switch>
OFF|ON
OFF: Fitting curve not displayed
ON: Fitting curve displayed