3.4 Device Message Details
3-71
3
Message Details
:CALCulate:LMARker:Y
Function
This command sets and displays the value of the level marker.
This command queries the value of level marker.
Syntax
:CALCulate:LMARker:Y
<switch>,<numeric>{DBM|MW|NW|PW|UW|W}
:CALCulate:LMARker:Y? 3|4
<switch>
Description
3
Specifes the level marker C.
4
Specifes the level marker D.
<numeric>
Log scale in absolute value
Range –190.000 to +50.000 dBm
Resolution 0.001
Linear scale in relative value
Range 0.001 pW to 1.200 W
Resolution 0.001
When the vertical axis shows relative values (dB, %), the level marker
value cannot be set or queried.
If a suffix is omitted, the unit will be dBm or W.
Response Data
<numeric>
Example of Use
:CALCulate:LMARker:Y 3,-20.55DBM
:CALCulate:LMARker:Y? 3
>–2.05500000E+001
:CALCulate:MARker:AOFF
Function
This command erases display of the wavelength, level, trace and delta
markers.