3.4 Device Message Details
3-173
3
:CALCulate:PARameter[:CATegory]:WFIL:SGWavelength
Function
This command sets and queries the channel wavelength detection
method at WDM filter application.
Syntax
:CALCulate:PARameter[:CATegory]:WFIL:SGWavelength
{<switch>[,<numeric>[DB]]}
:CALCulate:PARameter[:CATegory]:WFIL:SGWavelength?
Detection Method for
Channel Wavelength
Res. 0.1 dB
If a suffix is omitted,
If Parameter 2 is omitted, Channel Detection Cut Level is not changed.
Response Data
PEAK|THRESHOLD|RMS[,<numeric>]
<numeric>: Cut Level (dB)
+1.00000000E
–
001 to +5.00000000E+001
Example of Use
:CALCulate:PARameter:CATegory:WFIL:SGWavelength PEAK,25
:CALCulate:PARameter:CATegory:WFIL:SGWavelength?
>PEAK
:CALCulate:PARameter[:CATegory]:WFIL:SLICe
Function
This command sets and queries the slice level to detect the channel at
WDM filter application.
Syntax
:CALCulate:PARameter[:CATegory]:WFIL:SLICe <numeric>[DB]
:CALCulate:PARameter[:CATegory]:WFIL:SLICe?
Range 0.1 to 50.0 dB
Resolution 0.1 dB
If a suffix is omitted, the unit will be dB.