3.4 Device Message Details
3-241
3
[:SENSe][:WAVelength]:CENTer
Function
This command sets and queries the center wavelength.
Syntax
[:SENSe][:WAVelength]:CENTer <numeric>[<unit_m>]
[:SENSe][:WAVelength]:CENTer?
Range 600.00 to 1750.00 nm
Resolution 0.01 nm
If a suffix is omitted, the unit will be m.
Response Data
<numeric>
<numeric>: Center wavelength (m)
+6.00000000E–007 to +1.75000000E–006
Example of Use
:SENSe:WAVelength:CENTer 1545350PM
:SENSe:WAVelength:CENTer?
>+1.54535000E-006