Chapter 3 Message Details
3-136
:CALCulate:PARameter[:CATegory]:NF:MDIFf
Function
This command sets and queries the threshold value for detecting the
peak (channel) in the Optical AMP (WDM) application.
Syntax
:CALCulate:PARameter[:CATegory]:NF:MDIFf <numeric>[DB]
:CALCulate:PARameter[:CATegory]:NF:MDIFf?
<numeric>: Search Threshold
Range 0.01 to 10.00 dB
Resolution 0.01 dB
If a suffix is omitted, the unit will be dB.
Response Data
<numeric>
<numeric>: Search Threshold (dB)
+1.00000000E–002 to +1.00000000E+001
Example of Use
:CALCulate:PARameter:CATegory:NF:MDIFf 6
:CALCulate:PARameter:CATegory:NF:MDIFf?
>+6.00000000E+000
:CALCulate:PARameter[:CATegory]:NF:MSELect
Function
This command selects the saving destination for measurement data from
the Optical AMP (WDM) application.
This command queries the saving destination for measurement data
from the Optical AMP (WDM) application.
Syntax
:CALCulate:PARameter[:CATegory]:NF:MSELect <switch>
:CALCulate:PARameter[:CATegory]:NF:MSELect?
<switch>: Measurement data saving destination {PIN|POUT}
Response Data
<switch>
Example of Use
:CALCulate:PARameter:CATegory:NF:MSELect PIN