Chapter 3 Message Details
3-168
:CALCulate:PARameter[:CATegory]:WDM:TH
Function
This command sets and queries the cut level of the Signal Parameter
Threshold method at WDM application.
Syntax
:CALCulate:PARameter[:CATegory]:WDM:TH <numeric>[DB]
:CALCulate:PARameter[:CATegory]:WDM:TH?
Range 0.1 to 50.0 dB
Resolution 0.1 dB
If a suffix is omitted, the unit will be dB.
Response Data
<numeric>
<numeric>: Cut Level (dB)
+1.00000000E
001 to +5.00000000E+001
Example of Use
:CALCulate:PARameter:CATegory:WDM:TH 10
:CALCulate:PARameter:CATegory:WDM:TH?
>+1.00000000E+001