EasyManua.ls Logo

Anritsu MS9740A - Calculate:threshold; Calculate[1|2|3|4|5|6]:Tpower[:Data]

Anritsu MS9740A
320 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.4 Device Message Details
3-181
3
Message Details
:CALCulate:THReshold
Function
This command sets and queries the slice level for the Optical AMP
(WDM) application.
Syntax
:CALCulate:THReshold <numeric>[DB]
:CALCulate:THReshold?
Range 0.1 to 50.0 dB
Resolution 0.1 dB
If a suffix is omitted, the unit will be dB.
Response Data
<numeric>
<numeric>: Slice Level (dB)
+1.00000000E
001 to +5.00000000E+001
Example of Use
:CALCulate:THReshold 0.1
:CALCulate:THReshold?
>+1.00000000E-001
:CALCulate[1|2|3|4|5|6]:TPOWer[:DATA]?
Function
This command queries the value of integrated power analysis of
spectrum level.
Syntax
:CALCulate[1|2|3|4|5|6]:TPOWer[:DATA]?
Specify the Active trace (A to F) by specifying a number from 1 to 6.
When omitted, the Active trace is not changed.
Response Data
<numeric>
<numeric>: Value of integral analysis of spectrum level (dBm)
Example of Use
:CALCulate1:TPOWer:DATA?
>-1.11200277E+001