3.4 Device Message Details
3-85
3
:CALCulate:MARKer:PP:DATA?
Function
This command queries the Peak to Peak measured result of trace.
When the level scale is displayed in units of percentage (%), the result of
“Peak to Peak” calculation cannot be queried.
Syntax
CALCulate:MARKer:PP:DATA?
Response Data
<numeric>
<numeric>: Peak to Peak measurement result (dB|W)
When Peak to Peak Calculation is NOT set to On, the response data
for :CALCulate:MARKer:PP:DATA? command is –999.99.
:CALCulate:MARKer:PP[:STATe]
Function
This command sets and queries the Peak to Peak display of trace.
Syntax
CALCulate:MARKer:PP[:STATe] 0|1|OFF|ON
CALCulate:MARKer:PP[:STATe]?
0|OFF: Does not display Peak to Peak.
1|ON: Displays Peak to Peak.
Response Data
OFF|ON
Example of Use
CALCulate:MARKer:PP:STATe 1
CALCulate:MARKer:PP:STATe?
>1