3.4 Device Message Details
3-83
3
:CALCulate:MARKer[1|2|3|4]:MINimum:RIGHt
Function
This command detects next dip point on the right after current point and
moves trace marker to point.
Syntax
:CALCulate:MARKer[1|2|3|4]:MINimum:RIGHt
The process will not change regardless of whether a number [1|2|3|4] is
specified or omitted.
Example of Use
CALCulate:MARKer:MINimum:RIGHt
:CALCulate:MARKer[1|2|3|4]:PEXCursion[:PEAK]
Function
This command sets the Search Threshold for Peak/Dip Search to Manual
and sets its value.
This command queries the Search Threshold value for Peak/Dip Search.
Syntax
:CALCulate:MARKer[1|2|3|4]:PEXCursion[:PEAK]
<numeric>[DB]
:CALCulate:MARKer[1|2|3|4]:PEXCursion[:PEAK]?
The process will not change regardless of whether a number [1|2|3|4] is
specified or omitted.
<numeric>: Search Threshold
Range 0.01 to10.00 dB
Resolution 0.01 dB
If a suffix is omitted, the unit will be dB.
Response Data
<numeric>
<numeric>: Search Threshold(dB)
+1.00000000E–002 to +1.00000000E+001
Example of Use
:CALCulate:MARKer:PEXCursion:PEAK 2
:CALCulate:MARKer:PEXCursion:PEAK?
>+2.00000000E+000