3.4 Device Message Details
3-147
3
:CALCulate:PARameter[:CATegory]:POWer:OFFSet
Function
This command sets the level offset and moves the screen waveform by
the level offset amount.
This command queries the level offset.
Syntax
:CALCulate:PARameter[:CATegory]:POWer:OFFSet
<numeric>[DB]
:CALCulate:PARameter[:CATegory]:POWer:OFFSet?
Range –30.00 to 30.00 dB
Resolution 0.01 dB
If a suffix is omitted, the unit will be dB.
Response Data
<numeric>
<numeric>: Level offset value (dB)
–3.00000000E+001 to +3.00000000E+001
Example of Use
:CALCulate:PARameter:CATegory:POWer:OFFSet 2.6
:CALCulate:PARameter:CATegory:POWer:OFFSet?
>+2.60000000E+000