Chapter 3 Message Details
3-186
Example of Use
:CALibration:AUToalign:ENABle 1
:CALibration:AUToalign:ENABle?
>1
:CALibration:PRESet
Function
This command returns the optical alignment data and wavelength
calibration data to the factory defaults.
Syntax
:CALibration:PRESet
:CALibration:RESCal
Function
This command sets the actual resolution value to the initial value or
correction value.
This command queries the actual resolution calibration status.
When processing is completed, bit 4 of the end event status register
(ESR2) is set to 1.
Syntax
:CALibration:RESCal 0|1
:CALibration:RESCal?
0: Uses default resolution calibration value
1: Executes resolution calibration and calculates resolution
calibration value
Response Data
0|1|2|3
0: Uses default resolution calibration value
1: Resolution calibration finished normally
2: Calibrating resolution
3: Resolution calibration finished abnormally
Example of Use
:CALibration:RESCal 1
:CALibration:RESCal?
>2