Chapter 3 Message Details
3-238
[:SENSe]:CORRection:RVELocity:MEDium
Function
This command sets and queries whether to display the wavelength in air
or in vacuum.
Syntax
[:SENSe]:CORRection:RVELocity:MEDium 0|1|AIR|VACuum
[:SENSe]:CORRection:RVELocity:MEDium?
0|AIR: Value in air
1|VACuum: Value in vacuum
Response Data
AIR|VAC
Example of Use
:SENSe:CORRection:RVELocity:MEDium AIR
:SENSe:CORRection:RVELocity:MEDium?
>AIR
:SENSe:CORRection:WAVelength:SHIFt
Function
This command sets and queries the wavelength offset and moves the
waveform on the screen by the offset.
Syntax
:SENSe:CORRection:WAVelength:SHIFt <numeric>[<unit_m>]
:SENSe:CORRection:WAVelength:SHIFt?
Range –1.00 to 1.00 nm
Resolution 0.01 nm
If a suffix is omitted, the unit will be m.
Response Data
<numeric>
<numeric>: Wavelength offset value (m)
–1.00000000E
–
009 to +1.00000000E–009
Example of Use
:SENSe:CORRection:WAVelength:SHIFt -0.05NM
:SENSe:CORRection:WAVelength:SHIFt?
>-5.00000000E-011