Chapter 3 Message Details
3-222
:MMEMory:MOVE:GRAPhics
Function
This command moves the graphics file from drive D to drives E to Z .
The extension (bmp or png) of the moved file is specified at Copy
Settings.
If the specified device or file cannot be found and an error is generated,
the standard event status execution error bit becomes 1.
The copy source graphics file should be moved to the following folder of
the specified drive.
x:\Anritsu Corporation\Optical Spectrum Analyzer\User Data\Screenshot
This message is a system management command.
Syntax
:MMEMory:MOVE:GRAPhics <file_name>,<user_drive>
It is not necessary to add the extension to file_name.
Example of Use
:MMEMory:MOVE:GRAPhics 'LED_125M(025)',F
:MMEMory:MOVE:SYSinfo
Function
This command moves the system information file from drive D to drives
E to Z.
If the specified device or file cannot be found and an error is generated,
the standard event status execution error bit becomes 1.
The system information file should be moved to the following folder of the
specified drive.
x:\Anritsu Corporation\Optical Spectrum Analyzer\User Data\System Information
This message is a system management command.
Syntax
:MMEMory:MOVE:SYSinfo <file_name>,<user_drive>
It is not necessary to add the extension to file_name.
Example of Use
:MMEMory:MOVE:SYSinfo 'SystemInfo-20131205_001',F