Anritsu MS9740A - Display[:Window]:Trace:y1[:Scale]:Blevel

Anritsu MS9740A
320 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Message Details
3-204
:DISPlay[:WINDow]:TRACe:Y1[:SCALe]:BLEVel
Function
The command sets the level scale to the linear and sets the Linear Level
value.
This command queries the linear Level value.
Syntax
:DISPlay[:WINDow]:TRACe:Y1[:SCALe]:BLEVel
<numeric>[MW|NW|PW|UW|W]
:DISPlay[:WINDow]:TRACe:Y1[:SCALe]:BLEVel?
<numeric>:
The units for absolute value display are from 1 pW to 1 W as follows:
MW:mW, UW:
µ
W, NW:nW, PW:pW,W:W
If the units are omitted, W is assumed.
When the scale is displayed in relative values (%), the Linear Level
value cannot be set.
Response Data
<numeric>: Linear Level value (W)
Example of Use
:DISPlay:WINDow:TRACe:Y1:SCALe:BLEVel 50UW
:DISPlay:WINDow:TRACe:Y1:SCALe:BLEVel?
>+5.00000000E-005