EasyManua.ls Logo

Anritsu MS9740A - Calculate:parameter[:Category]:Pmd

Anritsu MS9740A
320 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Message Details
3-146
:CALCulate:PARameter[:CATegory]:PMD
Function
This command sets the parameters and performs PMD application
analysis.
This command queries the parameters for the PMD application.
Syntax
:CALCulate:PARameter[:CATegory]:PMD
<numeric>[DB],<switch>,[<numeric>]
:CALCulate:PARameter[:CATegory]:PMD?
No.
Parameter Type
Range
Description
1
<numeric>
0.01 to 1.00 dB
Res. 0.01 dB
Mode Coupling factor (dB)
If a suffix is omitted, the
unit will be dB.
2
<switch>
0|1
Selecting measurement
method
0: Auto 1: Manual
3
<numeric>
2 to 99
Res. 1
Peak Count
Measurement method 0: can
be omitted when Auto is set.
Response Data
<numeric>,<switch>,[<numeric>]
No.
Parameter Type
Description
1
<numeric>
Mode Coupling factor (dB)
2
<switch>
Selecting measurement method
0: Auto 1: Manual
3
<numeric>
Peak Count
Example of Use
:CALCulate:PARameter:CATegory:PMD 0.8,1,8
:CALCulate:PARameter:CATegory:PMD?
>+8.00000000E-001,1,8