Chapter 3 Message Details
3-140
Response Data
<switch>,<switch>,<numeric>,<numeric>,<numeric>,<switch>
,<switch>
No.
Parameter type
Description
1
<switch>
0:NF (S-ASE)
1:NF (Total)
2
<switch>
0: ISS Method (IEC)
1: ISS Method (Advanced)
2: Off
3
<numeric>
Pin Loss(Offset) (dB)
Loss correction factor for signal level
4
<numeric>
Pout Loss(Offset) (dB)
Loss correction factor for optical level output
5
<numeric>
NF Calibration
Correction factor for noise figure
6
<switch>
0: Actual Resolution (Measured)
1: Actual Resolution (Initial)
7
<switch>
OFF: Fitting curve not displayed
ON: Fitting curve displayed
Example of Use
:CALCulate:PARameter:CATegory:NF:PARameter
0,2,10,5,10,0,ON
:CALCulate:PARameter:CATegory:NF:PARameter?
>0,2,+1.00000000E+001,+5.00000000E+000,+1.00000000E+001,
0,ON