3.4 Device Message Details
3-149
3
Example of Use
:CALCulate:PARameter:CATegory:SWENvelope:TH1 10
:CALCulate:PARameter:CATegory:SWENvelope:TH1?
>+1.00000000E+001
:CALCulate:PARameter[:CATegory]:SWPKrms
Function
This command sets and queries slice level and factor K for the RMS
method.
Syntax
:CALCulate:PARameter[:CATegory]:SWPKrms
<numeric>[DB],<numeric>
:CALCulate:PARameter[:CATegory]:SWPKrms?
Parameter 1: Slice level
<numeric>:
Range 0.1 to 50.0 dB
Resolution 0.1 dB
If a suffix is omitted, the unit will be dB.
Parameter 2:
<numeric>: Standard deviation factor k
Range 1.00 to 10.00
Resolution 0.01
Response Data
<numeric>,<numeric>
Parameter 1:
<numeric>: Slice level (dB)
+1.00000000E–001 to +5.00000000E+001
Parameter 2:
<numeric>: Standard deviation factor k
+1.00000000E+000 to +1.00000000E+001
Example of Use
:CALCulate:PARameter:CATegory:SWPKrms 6,2.35
:CALCulate:PARameter:CATegory:SWPKrms?
>+6.00000000E+000,+2.35000000E+000