Chapter 2 Before Use
2-22
End Event Register
The meaning of each bit of the end event status register is listed in the
table below.
Table 2.6.3-2 Meaning of End Event Status Register
Bit Explanation
7
Not used; always 0
6
Not used; always 0
5
Not used; always 0
4
End Execution 1
Becomes 1 when calibrating wavelength of resolution
and adjusting optical system operations completed.
3
End Execution 2
Becomes 1 when sweep averaging or measuring with
power monitor operations completed.
2
Not used; always 0
1
Sweep end
Become 1 when sweeping completed.
0
Measurement end
Becomes 1 when one of the following commands has
been processed: Auto Measurement, analysis using
Analysis function, Peak/Dip Search processing,
analysis using Application function.
To execute multiple commands,
send :
STATus:EVENt:CONDition?
for each command
to query end event register.