CAPINTEC, INC. CAPTUS
®
4000e
10-40 LAB TESTS November 16
Note: If the background has recently been counted, then the screen shown in Figure
10-57 will appear. To skip counting the background again and proceed to step 5,
select the Yes button or press the Y key. To re-measure the background, select
the No button or press the N key. The screen shown in Figure 10-58 will appear.
Figure 10-57
Figure 10-58
To begin counting, verify that there are no radioactive sources nearby and select the OK
button. To abort the measurement, select the Cancel button and the screen will return to
that shown in Figure 10-55. To exit to main Cr51 Blood Volume test screen, select the
Exit button or press Alt+E. To begin the test again, select the Start Procedure button or
press Alt+S.
5. The screen appears as shown in Figure 10-59, prompting you to measure the whole
blood standard.
Figure 10-59
To begin counting, place the standard in the Well and select the OK button. To quit
counting, select the Cancel button. This will provide a prompt to abort the patient test, as
shown in Figure 10-60.