EasyManua.ls Logo

Dräger Isolette C2000 - Page 199

Dräger Isolette C2000
524 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Theory of Operation
Chapter 3: Theory of Operation
Page 3 - 38 Isolette® Infant Incubator (Model C2000) Service Manual
Table 3-2. ID Character
Table 3-3. Mode Bit Flags
For example: If “Humidity on” and “Baby Mode configuration” are
selected, the character is 05.
Character Description
<space> Normal mode
1 Special/Test Mode is in effect (data may be
invalid)
Bit Description
01 Humidity on
02 Oxygen on
04 Baby Mode configuration
08 0.5 C baby alarm limit
10 Reserved
20 Reserved
40 Reserved
80 Reserved

Table of Contents

Related product manuals