Theory of Operation
Chapter 3: Theory of Operation
Page 3 - 38 Isolette® Infant Incubator (Model C2000) Service Manual
Table 3-2. ID Character
Table 3-3. Mode Bit Flags
For example: If “Humidity on” and “Baby Mode configuration” are
selected, the character is 05.
Character Description
<space> Normal mode
1 Special/Test Mode is in effect (data may be
invalid)
Bit Description
01 Humidity on
02 Oxygen on
04 Baby Mode configuration
08 0.5 C baby alarm limit
10 Reserved
20 Reserved
40 Reserved
80 Reserved