Theory of Operation
Chapter 3: Theory of Operation
Isolette® Infant Incubator (Model C2000) Service Manual Page 3 - 39
3
Table 3-4. Alarm Bit Flags
For example: If the air temperature and oxygen are low and Procedural
Silence is initiated, such as when an access door is open, the 10
character value equals 000000405.
Certain fields, such as air temperature, have an implied decimal point.
The decimal point does not physically appear in the data stream.
Bit Description
0000000001 Low control temperature
0000000002 High control temperature
0000000004 Low oxygen
0000000008 High oxygen
0000000010 High temperature cut-out
0000000020 Skin 1—probe failure
0000000040 Skin probe—disconnect
0000000080 Oxygen calibration required
0000000100 Sensor out of position
0000000200 Water level low
0000000400 Procedural Silence
0000000800 Motor failed
0000001000 Low air flow
0000002000 Heater failed
0000004000 EEPROM failed
0000008000 Sensor module failure
0000010000 Controller failure 1
0000020000 Controller failure 2
0000040000 Controller failure 3
0000080000 Controller failure 4
0000100000 Air probe failed
0000200000 Oxygen cell different
0000400000 Scale disconnect
0000800000 Too much weight
0001000000 Scale failed

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Isolette C2000 and is the answer not in the manual?

Dräger Isolette C2000 Specifications

General IconGeneral
BrandDräger
ModelIsolette C2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals