6.1 Installation and Set-up
Chapter 6: General Procedures
Page 6 - 18 Isolette® Infant Incubator (Model C2000) Service Manual
Table 6-4. Set-up Options, Parameters, and Modes
4. Press the Up arrow key (M) to scroll to a language, raise the air set
temperature, or select the following:
• YES (activate)
• 100%
•1.0°C
• lb (pounds)
5. Press the Down arrow key (N) to scroll to a language, lower the air
set temperature, or select the following:
• NO (deactivate)
• 21%
•0.5°C
• kg (kilograms)
6. To exit, press the Silence/Reset key (K).
Display Selection key
selections
Up Arrow key
selections
Down Arrow key
selections
Humidity option Yes No
Oxygen option Yes No
Oxygen calibration
level
100% 21%
Skin temperature
alarm limit
1.0 °C 0.5 °C
Skin Mode Yes No
Language English Italian
French Spanish
German German
Spanish French
Italian English
Weight unit lb kg
Air set temperature 30 °C to 37 °C 37 °C to 30 °C
Altitude < 10000' (3048 m) > 10000' (3048 m)