Function Checks
Chapter 2: Troubleshooting Procedures
Page 2 - 8 Isolette® Infant Incubator (Model C2000) Service Manual
display on Display #1. Go to “System Failure Messages”
on page 2-33.
3. The Trend display is free from any alarm or system prompt
messages.
Yes No
↓→Go to “Alarms” on page 2-25.
4. Perform the following:
a. Allow the unit to operate for at least 3 minutes.
b. Check the Power Failure alarm by unplugging the power cord
from its power source.
The Power Failure alarm sounds, and the Power Fail light-emitting
diode (LED) on the controller actuates.
Yes No
↓→Go to RAP 2.1 on page 2-51.
5. Perform the following:
a. Plug the power cord into an appropriate power source. Ensure
that the power cord is fully plugged into the wall, the pedestal
stand, and the incubator receptacles.
b. S et the air set temperature to 95 F (35 C). Allow the unit to
operate while checking the hood/shell.
The unit has a variable height adjustable (VHA) pedestal stand (A)
(see figure 2-1 on page 2-9).
Yes No
↓→Go to step 8.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Isolette C2000 and is the answer not in the manual?

Dräger Isolette C2000 Specifications

General IconGeneral
BrandDräger
ModelIsolette C2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals