Function Checks
Chapter 2: Troubleshooting Procedures
Page 2 - 10 Isolette® Infant Incubator (Model C2000) Service Manual
6. Perform the following to check the operation of the VHA
pedestal/stand (A):
a. Using your foot, press the Up arrow on the front Up/Down switch
(C) to raise the VHA pedestal/stand (A) to the maximum height.
b. Press and hold the Down arrow on the Up/Down switch (C) to
lower the VHA pedestal/stand (A) to the minimum height.
c. Repeat using the rear Up/Down switch (C).
The VHA pedestal/stand (A) operates smoothly and adjusts to the
desired height.
Yes No
↓→Go to RAP 2.2 on page 2-52.
7. The unit is equipped with a non-locking hood assembly (D).
Yes No
↓→Go to step 9.
8. Perform the following to check the operation of the non-locking hood
assembly (D) (see figure 2-1 on page 2-9):
a. Disconnect the weighing scale cable (E) from the sensor module
(F).
b. Slowly tilt the hood assembly (D) back until it comes to rest.
c. Close the hood assembly (D).
d. Connect the weighing scale cable (E) to the sensor module (F).
The hood assembly (D) opens and closes correctly.
Yes No
↓→Go to RAP 2.6 on page 2-60.
9. The hood assembly (D) is equipped with a pivot/hood lock assembly
bracket (G).
Yes No
↓→Go to step 11.
10.On the hood assembly (D) equipped with a pivot/hood lock
assembly bracket (G), perform the following:
a. Disconnect the weighing scale cable (E) from the sensor module
(F).
b. Slowly tilt the hood assembly (D) back until it locks in place.