Alarms
Chapter 2: Troubleshooting Procedures
Page 2 - 32 Isolette® Infant Incubator (Model C2000) Service Manual
Table 2-7. Silence/Reset Key Functions
Alarm Silence/Reset Key Function
Low Air/Skin Temperature Silence the alarm for a period of 15 minutes.
High Air/Skin Temperature Silence the alarm for a period of 15 minutes.
Humidity Low Silence the alarm for a period of 15 minutes.
Power Failure Silence the alarm for a period of 15 minutes.
Skin 1 Probe Failure Silence the alarm for a period of 5 minutes.
Skin Probe Disconnect Silence the alarm for a period of 5 minutes.
High Temperature Cut Out Silence the alarm for a period of 5 minutes.
Oxygen High % Silence the alarm for a period of 4 minutes.
Oxygen Low % Silence the alarm for a period of 4 minutes.
Oxygen Calibration Required Silence the alarm for a period of 4 minutes.
Oxygen Cell Different Silence the alarm for a period of 4 minutes.
High Temperature Cut Out Resets the alarm, or cancels the alarm if the alarm
condition no longer exists.
Sensor Module Failure Resets the alarm, or cancels the alarm if the alarm
condition no longer exists.
Power Fail Resets the alarm, or cancels the alarm if the alarm
condition no longer exists.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Isolette C2000 and is the answer not in the manual?

Dräger Isolette C2000 Specifications

General IconGeneral
BrandDräger
ModelIsolette C2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals