EasyManua.ls Logo

GE Vivid S6 N - Section 3-5 Completing the Hardware Installation

GE Vivid S6 N
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
3-26 Section 3-4 - Preparing for Installation
PRELIMINARY
Section 3-5
Completing the Hardware Installation
The hardware installation procedures described in this section provide instructions and information
regarding connecting a printer, peripherals and probes, etc.
NOTE: If required, it is possible to disable access to all Vivid S5 N / Vivid S6 N scanner USB ports (including
the CD/DVD drive) in order to limit system vulnerability. This is implemented from the System
Configuration window, via the Service tab.
Activate the USB External Media disabled checkbox (as shown in Figure 3-20 below), then press the
F2 or Esc key to save the changes.
NOTE: After disabling all Vivid S5 N / Vivid S6 N scanner USB ports, all peripherals that are used by connecting
to a USB port (and also the CD/DVD drive) will not be accessible until such time as access to USB
External Media is re-enabled - i.e., the USB External Media disabled checkbox (shown above) is
unchecked.
Figure 3-20 Disabling USB External Media

Table of Contents

Other manuals for GE Vivid S6 N

Related product manuals