8-52 Section 8-3 - Control Console Components Replacement Procedures
PRELIMINARY
8-3-6 Probe Shelf Replacement Procedure
8-3-6-1 Tools
Use the appropriate screwdriver as indicated in the Probe Shelf Replacement procedure.
8-3-6-2 Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
8-3-6-3 Probe Shelf Removal Procedure
1) Remove the Keyboard Assembly as described in the Keyboard Assembly Removal Procedure” on
page 8-39.
2) Working from below, and while carefully supporting the Probe Shelf Holder, remove the 4 retaining
screws - two on either side - as seen in Figure 8-65.
3.) Remove the Probe Shelf.
8-3-6-4 Probe Shelf Installation Procedure
1.) Return the Probe Shelf to its location on the Keyboard Interface Assembly and secure it with the
4 retaining screws (previously removed).
2.) Install the Keyboard Assembly as described in the “Keyboard Assembly Installation Procedure” on
page 8-42.
3) Lower the console to the maximum down position.
4.) Turn ON power to the system.
5.) Check for proper up/down and side-to-side movement of the operator panel and monitor, as
described in “Mechanical Functional Tests” on page 4-35.
6.) Proceed to perform the following functionality tests:
- “Grounding Continuity” on page 10-23
- “Chassis Current Leakage Test” on page 10-24
Figure 8-65 Probe Shelf - Viewed from Below indicating Retaining Screws
Retaining Screw
Retaining Screw