ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 3 - System Setup 3-129
PRELIMINARY
3-7-6 Query/Retrieve (Q/R) Setup
3-7-6-1 Overview
The Query/Retrieve function makes it possible to search for and retrieve DICOM data from a DICOM
server for further analysis on the Vivid S5 N or Vivid S6 N.
3-7-6-2 Query/Retrieve Setup on the Vivid S5 N or Vivid S6 N
1.) Press CONFIG
(F2) and log on as Adm, as described in ‘Log On to the System as ADM” on page 4-
10.
2.) Select Connectivity.
3.) Select Dataflow.
4.) Select Query Retrieve from the Name pull-down menu - see Figure 3-152 on page 3-130.
The selected dataflow (Query/Retrieve), is displayed as shown in Figure 3-151.
Figure 3-150 Select Query Retrieve
NAME
PULLDOWN
MENU
QUERY RETRIEVE