ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
2-10 Section 2-3 - Facility Needs
PRELIMINARY
2-3-4 Networking Pre-Installation Requirements
2-3-4-1 Stand-alone Unit (without Network Connection)
None.
2-3-4-2 Unit Connected to Hospital’s Network
Supported networks:
• Wireless LAN
2-3-4-3 Purpose of the DICOM Network Function
DICOM services provide the operator with clinically useful features for moving images and patient
information over a hospital network. Examples of DICOM services include the transfer of images to
workstations for viewing or transferring images to remote printers. As an added benefit, transferring
images in this manner frees up the on-board monitor and peripherals, enabling viewing to be done while
scanning continues. With DICOM, images can be archived, stored, and retrieved faster, easier, and at
a lower cost.
2-3-4-4 DICOM Option Pre-Installation Requirements
To configure the Vivid S5 N or Vivid S6 N ultrasound unit to work with other network connections, the
network administrator must provide the required information, which should include the following:
• Details: DICOM network details for the Vivid S5 N or Vivid S6 N unit,
including the host name, local port, IP address, AE title and net
mask.
• Routing Information: IP addresses for the default gateway and other routers in use at
the site.
• DICOM Application Information:Details of the DICOM devices in use at the site, including the
DICOM host name, AE title and IP addresses.