EasyManua.ls Logo

GE Vivid S6 N - 4-2-4 Log On to the System as ADM

GE Vivid S6 N
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
4-10 Section 4-2 - General Procedures
PRELIMINARY
4-2-4 Log On to the System as ADM
Table 4-2 Logging on to the System as ADM
Step Task Expected Result(s)
1. Press Config on the alphanumeric keyboard.
If not already logged on, the Operator Login window is displayed.
Note: The default password is ulsadm
2. Select the down-arrow to the right of the Operator field to view
the Operator pull down list.
From the factory, the following two users are defined without a
password:
- adm
- usr
It is possible for the administrator (adm) to establish unique
passwords for each user, but do not do this now.
3. Select the user adm.
- From the factory, the password has been left blank, so you may
try to log on without using any password.
- If the log-on fails, you may try to use the password service.
- As a last resort, contact the hospital’s Network Administrator
(or person responsible for the ultrasound unit), and ask for the
password.
The Emergency button stores data only from current patient
examination.
4. Select Log on.
- If this is the first time the unit is turned on, the “IMAGING AND
ANALYSIS - GLOBAL LEVEL” window is displayed.
- If another screen was displayed earlier, before logging out or turning the
unit OFF, that screen will be displayed.

Table of Contents

Other manuals for GE Vivid S6 N

Related product manuals