EasyManua.ls Logo

GE Vivid S6 N - 8-9-3 MOD Drive Replacement Procedure; 8-9-3-1 Tools; 8-9-3-2 Time Required; 8-9-3-3 Preparations

GE Vivid S6 N
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-240 Section 8-8 - Software Loading
PRELIMINARY
8-9-3 MOD Drive Replacement Procedure
NOTE: This section describes removal and replacement procedures only.
For initial installation instructions, refer to the information provided in Chapter 3 - System Setup. See
the MOD Drive section, page 3-67.
8-9-3-1 Tools
None.
8-9-3-2 Time Required
30 mins
8-9-3-3 Preparations
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
8-9-3-4 MOD Drive Removal Procedure
1.) Disconnect the USB cable from the scanner rear connectors panel, and from the USB port at the
rear of the MOD Drive.
2.) Disconnect the power cable from the power IN socket at rear of the MOD Drive, and from the
peripheral power outlet at the rear of the scanner.
3.) Remove the MOD Drive and place on a flat surface (together with the two cables removed).
8-9-3-5 MOD Drive Installation Procedure
1.) Place the MOD Drive either on the tray at the rear of the Vivid S5 N or Vivid S6 N scanner, or on a
flat, stable surface close to the scanner.
2.) Using the power cable (provided with the MOD Drive), connect one end to the power IN socket at
rear of the MOD Drive and the other to one of the peripheral power outlets at the rear of the scanner.
Figure 8-275 MOD Drive