ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-6 Section 8-2 - Cover Replacement Procedures
PRELIMINARY
8-2-3 Left Side Cover Replacement Procedure
NOTE: The left side cover is fastened in position on the side of the system by 4 ball and snap-lock securing clips.
This cover contains an air filter, held in position by securing clips.
For instructions on replacing the air filter only, see “Air Filter Replacement Procedure” on page 8-7.
8-2-3-1 Tools
None.
8-2-3-2 Preparations
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
8-2-3-3 Left Side Cover Removal Procedure
1) Gripping the recess located near the bottom of the cover (as shown in Figure 8-4), pull the bottom
of the cover out towards you to release it from the lower securing clip.
Figure 8-4 Removing the Left Side Cover
2) Pull the top of the cover out towards you, releasing it from the upper securing clip.
3) Remove the left side cover (together with the attached air filter) from the system.
Figure 8-5 Left Side Cover Removed
Grip cover here