ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
4-2 Section 4-2 - General Procedures
PRELIMINARY
Section 4-2
General Procedures
4-2-1 Overview
Since some general procedures are used more than others, for convenience, all the procedures
described in this section are listed below:
• Power ON/Boot-up section, page 4-2
• Power Shut Down section, page 4-7
• Log On to the System as ADM section, page 4-10
• Using Removable Media section, page 4-12
• Labeling Removable Media section, page 4-14
• Formatting Removable Media section, page 4-14
• DICOM Repair of Removable Media section, page 4-15
• Archiving and Restoring System Configuration section, page 4-15
4-2-2 Power ON/Boot-up
4-2-2-1 Connecting Mains Power to the Vivid S5 N or Vivid S6 N Ultrasound Unit
To avoid the risk of fire, power to the system must be supplied from a properly
rated outlet. It is recommended to use a dedicated power outlet. The power plug
should not, under any circumstances, be altered to a configuration rated less
than that specified for the current.
DO NOT use an extension cord or adaptor plug. Refer to the Electrical
Requirements section, on page 2-4 for more details.