Chapter 3 - System Setup 3-133
PRELIMINARY
3-7-6-3 Query/Retrieve Verification
Follow the Steps Below to do a First Test (Ping) of the Connection
1.) Select the Smiley button to Ping the server.
2.) If the network connection to the server is OK, it will be illustrated by a smiling Smiley .
A sad Smiley indicates that the network connection is failing.
Typical cause: Network cable not connected.
3.) When ready, click OK to close the DICOM Query/Retrieve properties and save changes.
Follow the Steps Below to Test the DICOM Query/Retrieve Workflow
NOTE: This check uses both Ping and DICOM Ping.
Figure 3-155 DICOM Query/Retrieve Properties
SMILEY
OK
VERIFICATION
CHECK BOX