GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-250 Section 8-8 - Software Loading
PRELIMINARY
8-9-6 Peripherals Console Replacement Procedure
8-9-6-1 Tools
Use the appropriate screwdrivers as indicated in the Peripherals Console replacement procedure.
8-9-6-2 Time Required
30 minutes
8-9-6-3 Preparations
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
8-9-6-4 Peripherals Console Removal Procedure
1) Remove the Vivid S5 N or Vivid S6 N left and right side covers, as described in the “Left Side Cover
Removal Procedure” on page 8-6 and “Right Side Cover Removal Procedure” on page 8-5.
2) Remove the Vivid S5 N or Vivid S6 N front cover, as described in the “Front Cover Removal
Procedure” on page 8-17.
3) Remove the Vivid S5 N or Vivid S6 N left and right rear covers, as described in the “Left Rear Cover
Removal Procedure” on page 8-13 and “Right Rear Cover Removal Procedure” on page 8-10
4) Remove the DVD/CD-RW drive, as described in the “DVD/CD-RW Drive Removal Procedure” on
page 8-236.
5) Remove the B&W printer, as described in the “Black and White Printer Removal Procedure” on
page 8-247.
6) Loosen the 4 retaining screws that secure the peripherals console in position, as shown in
Figure 8-286.
7. Gently slide the peripherals console forward, lift it slightly and then remove it.
Figure 8-286 Location of Retaining Screws