GE Vivid S6 N - 8-2-3-4 Left Side Cover Installation Procedure; 8-2-3-4-1 Air Filter Replacement Procedure

GE Vivid S6 N
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-7
PRELIMINARY
8-2-3 Left Side Cover Replacement Procedure (cont’d)
8-2-3-4 Left Side Cover Installation Procedure
1) Return the left side cover (together with the attached air filter) to the correct position on the left side
of the system, carefully aligning it with the securing clips.
2) Holding the recess with one hand, push the top of the cover upwards and inwards with the other
hand, until clicks into place.
3) Push the bottom of the cover upwards and inwards, until it clicks into place.
4.) Proceed to perform the following functionality tests:
- “Grounding Continuity” on page 10-23
- “Chassis Current Leakage Test” on page 10-24
- “Temperatures Test” on page 7-15
- “Fans Speed Test” on page 7-15
- “Fans Speed Setting Test” on page 7-15
8-2-3-4-1 Air Filter Replacement Procedure
NOTE: Commencing with Vivid S5 N and Vivid S6 N ultrasound scanners version BT12, systems are supplied
without an air filter. Therefore, for all BT versions, it is no longer necessary to perform the procedure
described below.
1) Remove the left side cover from the system, as described in “Left Side Cover Replacement
Procedure” on page 8-6.
2) Lay the left side cover face-down on a flat, clean surface.
3.) Undo the securing clip that holds one end of the filter in position on the cover, as shown in
Figure 8-6.
Figure 8-6 Releasing Air Filter Securing Clip
Release clip here