3-3-2 Verifying the Shipping Carton Contents
After unpacking, it is important to verify that all items ordered by the customer have been received.
Compare all items listed on the packing slip (shipping consignment note) with those received and report
any items that are missing, back-ordered, or damaged, to your GE HEALTHCARE sales representative.
The contact address is shown in ‘Contact Information” on page 1-27.
3-3-3 Physical Inspection
3-3-3-1 System Voltage Settings
Verify that the Vivid S5 N or Vivid S6 N ultrasound AC is set to the correct voltage.
The Voltage settings are:
220-240V AC - Europe, all Asia excluding Japan and Taiwan.
100-120V AC - Americas, Japan and Taiwan.
NOTE: Applicable to Vivid S5 N / Vivid S6 N systems with AC Distribution Box Type B or Type A.
3-3-4 EMI Protection
The Vivid S5 N or Vivid S6 N Ultrasound Unit has been designed to minimize the effects of Electro
Magnetic Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect
the system from image artifacts caused by this interference. For this reason, it is imperative that all
covers and hardware are installed and secured before the unit is put into operation.
NOTICE
IT IS RECOMMENDED TO KEEP AND STORE THE SHIPPING CARTON AND ALL OTHER
PACKING MATERIALS (INCLUDING THE SUPPORT FOAMS, ANTI-STATIC PLASTIC COVER,
ETC.), IN CASE THE UNIT HAS TO BE MOVED TO A DIFFERENT LOCATION IN THE FUTURE.
FOR WARRANTY PURPOSES, STORAGE OF THE ABOVE IS REQUIRED FOR ONE YEAR FROM
DATE OF PURCHASE.
WARNINGWARNING
SETTING THE Vivid S5 N or Vivid S6 N ULTRASOUND COMPONENTS TO THE WRONG
VOLTAGE SETTING WILL MOST LIKELY DESTROY THE EQUIPMENT.