ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-95
PRELIMINARY
Section 8-5
Cable Replacement Procedures
8-5-1 Keyboard and Monitor Cable Replacement Procedure
The Vivid™ S5 N and Vivid™ S6 N system’s communication and power cables extend from the LCD
monitor downwards along the Monitor Arm Assembly, to the Keyboard Interface Assembly, and finally
to the Distribution Interface Board (DIB). The replacement procedures are carried out in stages, as
described in the sub-sections below.
The following stages show the order in which cables are removed/replaced:
LCD monitor
Keyboard assembly
Upper/Lower Arm Assembly
Keyboard Interface Assembly
•DIB
8-5-1-1 Tools
Use the appropriate flat and Phillips screwdrivers, Allen Key, appropriate flat spanner, and diagonal side
cutter as indicated in the Keyboard or Monitor Cable replacement procedure.
8-5-1-2 Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
8-5-1-3 Keyboard and Monitor Cable Removal Procedure
1.) Remove the Monitor, as described in “Monitor Removal Procedure” on page 8-26.
2.) Remove the three Allen screws from the LDC metal mounting plate, as show in Figure 8-108.
8-5-1-3-1 Removing Cables from the Monitor Arm Assembly
Figure 8-108 Screws Securing Mounting Plate on the Support Column
Mounting Plate