ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-167
PRELIMINARY
8-6-12 Peripherals AC Outlet Connectors Panel Replacement Procedure
8-6-12-1 Tools
Use the appropriate flat and Phillips-type screwdrivers, and a suitably-sized T-Handle Hex Wrench, as
indicated in the Peripherals AC Outlet Connectors Panel replacement procedure.
8-6-12-2 Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
8-6-12-3 Peripherals AC Outlet Connectors Panel Removal Procedure
1) Disconnect any peripherals currently connected to the system.
2) Remove the system left and right side covers, as described in the “Left Side Cover Removal
Procedure” on page 8-6 and “Right Side Cover Removal Procedure” on page 8-5.
3) Remove the system front cover, as described in the “Front Cover Removal Procedure” on page 8-17.
4) Remove the system bottom rear cover, as described in the “Bottom Rear Cover Removal
Procedure” on page 8-23.
5) Loosen and remove the two screws that secure the Peripherals AC Outlet Connectors Panel, as
shown in Figure 8-208.
Figure 8-208 Peripherals AC Outlet Connectors Panel Securing Screws
Accessing the rear screw