ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 4 - Functional Checks 4-15
PRELIMINARY
4-2-8 DICOM Repair of Removable Media
1.) Insert the removable media to be verified.
2.) Select the pre-configured removable media from the Config > Connectivity > Tools pop-up menu.
3.) Press Repair DICOM Dir.
The unit starts to verify the DICOM directory on the selected media (if available), or (if not available)
creates a DICOM directory.
4-2-9 Archiving and Restoring System Configuration
NOTE: Always save presets before any software reload. This action is necessary to ensure that the presets
loaded after the software reload are as up–to–date as possible.
All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an
MO disk for reloading on the system.
Presets should not be saved on the same MO disk as images, because the Archive Menu lists the
images but not the presets stored on a MO disk.