ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
5-10 Section 5-3 - Front End Unit
PRELIMINARY
5-3-3 RFI_AM Board
The RFI_AM Board combines the following functionalities:
Front End Control
Radio Frequency & Tissue Processing
PCI Bus Interface
5-3-3-1 Tasks Performed by the RFI_AM Board
System Master Clock generator
Transmitter Control
ATGC Control
Receive Focusing Control
Analog Test Signal Generator
Tx voltage Enable/Disable
Probe Sensing
Control of TR32, PIB or PSB Boards
Ultrasound data acquisition and processing
Figure 5-8 RFI_AM Block Diagram
5-3-3-2 Watchdog and Power ON Reset
For safety reasons, a HW watchdog stops ultrasound transmit and resets the DSP if the DSP stops
toggling the watchdog inputs.
Once enabled, it is not possible to disable the watchdog unless the board is reset, or power is cycled.
5-3-3-3 Boot-up of DSP and Firmware
When the power is turned ON, the DSP is reset, and all FPGAs are blank. The DSP is connected to PCI
directly, and can only boot up with help from the Host.