Loading...
GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-20 Section 8-2 - Cover Replacement Procedures
PRELIMINARY
8-2-7 Basket Replacement Procedure
8-2-7-1 Tools
None
8-2-7-2 Preparations
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
8-2-7-3 Basket Removal Procedure
1) Remove all contents from the basket.
2) Working from the rear of the system, hold the basket with both hands and carefully lift it upwards to
release it from the two snap-lock securing clips - one on either side of the keyboard interface
column - see Figure 8-21.
3) Remove the basket.
8-2-7-4 Basket Installation Procedure
1) Working from the rear of the system, return the basket to its original position on the keyboard
interface column (see close-up view in Figure 8-21, above.
2) Holding the basket with both hands, carefully push it downwards to secure it on the two snap-lock
securing clips - one on either side of the column - refer to Figure 8-21.
3.) Proceed to perform the following functionality tests:
- “Monitor Movement Test” on page 4-36
- “Operator Panel Movement Test” on page 4-35
Figure 8-21 Position of Basket Securing Clips
CAUTION
The Basket has a maximum carrying capacity of 4 Kgs.
DO NOT EXCEED THIS LIMIT!
Keyboard Interface Column
Securing Clips
Close-up View of Basket In Position