GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 3 - System Setup 3-61
PRELIMINARY
In addition, a dedicated power source and cable is provided within the dedicated B&W printer
compartment.
The B/W Printer should be connected once the system is powered OFF.
All software drivers for the B/W Printer are pre-configured and installed designated to be used with the
specific B/W Printer supplied by GE HEALTHCARE.
3-5-1-6-1 Configuring the Sony UP-D897 B/W Video Printer
1.) Click the Tcpip tab on the Connectivity screen (see Figure 3-28 on page 3-33).
2.) Select Advanced Settings - see Figure 3-71.
3.) The Network Connection screen opens (Figure 3-72). Click the “Up” button to open the Windows
Control Panel.
DO NOT ATTEMPT TO USE A DIFFERENT TYPE OF B/W PRINTER (BRAND OR MODEL)
OTHER THAN THE B/W PRINTER PROVIDED BY GE. THE ULTRASOUND SYSTEM IS AN
EXTREMELY SENSITIVE AND COMPLEX MEDICAL SYSTEM.
ANY UNAUTHORIZED PERIPHERALS MAY CAUSE SYSTEM FAILURE OR DAMAGE!
Figure 3-71 Tcpip Screen
Figure 3-72 Network Connections showing highlighted “Up One Level” button