GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-184 Section 8-6 - Lower Section Components Replacement Procedures
PRELIMINARY
8-6-17 AC Distribution Box (Type A) Replacement Procedure
8-6-17-1 Tools
Use the appropriate flat and Phillips-type screwdrivers as indicated in the AC Distribution Box
replacement procedure.
8-6-17-2 Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
8-6-17-3 AC Distribution Box Removal Procedure
1.) Remove the bottom rear cover from the system, as described in “Bottom Rear Cover Removal
Procedure” on page 8-23.
As shown in Figure 8-232 below, pull the protective cover outwards and downwards to reveal the
AC Input Box assembly - refer to Figure 8-233.
Figure 8-232 Bottom Rear Cover - View from Side