GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-183
PRELIMINARY
3.) Return the two supporting screws (previously removed) to their positions and fasten them securely
(shown in Figure 8-229).
4.) Connect the Printer power cable (PN5448872) to the printer and attach it to the system chassis
using the retaining clips previously installed.
5.) Reconnect the two ground leads - as illustrated (on right) in Figure 8-231. Refer to note below.
Note: When reconnecting the ground leads, there are two spring washers that should be installed in the
order shown in Figure 8-231. In addition, place the ground lead connector with the flat surface
facing towards the system to prevent damage to the cable holder and the cable.
6.) Return the bottom rear cover to the system chassis, as described in “Bottom Rear Cover Installation
Procedure” on page 8-23.
7.) On the Left Rear Cover, insert the plastic AC Outlet Cover to blank the two AC Outlet ports
(previously used for connection of peripherals) - refer to Figure 8-227 on page 8-180.
8.) Adhere the plain white label onto the Left Rear Cover to occlude the power details (Max. 250VA
Total), now no longer relevant - as shown in Figure 8-227 on page 8-180.
9.) Turn ON power to the system.
10.)Proceed to perform the following functionality tests:
- “Grounding Continuity” on page 10-23
- “Chassis Current Leakage Test” on page 10-24
- “Calibration” on page 7-12
- “Full System Test” on page 7-11
Figure 8-231 Ground Lead Showing Spring Washers and Cable Holder
Ground Lead
Ground Lead
Ground Lead Fastened to New AC Distribution Box